- European Journal of Technique
- Volume:11 Issue:2
- Measurement of the Number of Photoelectrons per MeV for Neutron Detectors
Measurement of the Number of Photoelectrons per MeV for Neutron Detectors
Authors : Emre İREN, Ferhat ÖZOK, Mustafa Nizametin ERDURAN
Pages : 136-139
Doi:10.36222/ejt.950944
View : 34 | Download : 4
Publication Date : 2021-12-30
Article Type : Research Paper
Abstract :The light gains of a detector system insert ignore into journalissuearticles values(PMT and scintillator); is characterized by the number of photoelectrons produced per unit of energy insert ignore into journalissuearticles values(usually MeV);. In this study, the results of light gain tests for 10 Hamamatsu R11833 model photomultiplier tubes insert ignore into journalissuearticles values(PMT); are shown. Various radioactive sources were used in conjunction with the EJ309 liquid scintillator to determine the light gain. The number of photoelectrons produced per unit energy was determined by comparing the size of the signal generated by the photons produced by PMT during the scintillation process using radioactive sources and the size of the signal generated when a single photo electron insert ignore into journalissuearticles values(SPE); was produced by the PMT. The test results show that the performances of the PMTs are sufficient for neutron detector arrays used in experiments with radioactive ion beams. The supply voltage of the photomultiplier tubes must be changed to achieve similar light gains.Keywords : Photomultiplier Tube, Single Photoelectron, Liquid Scintillator, Photoelectron Yield, Neutron Detection