- Investigation of Surface Dose Accuracy of Two Dose Calculation Algorithms Using Thermoluminescent Do...
Investigation of Surface Dose Accuracy of Two Dose Calculation Algorithms Using Thermoluminescent Dosimeters
Authors : Osman Vefa GÜL
Pages : 353-360
Doi:10.54287/gujsa.1347041
View : 47 | Download : 33
Publication Date : 2023-09-29
Article Type : Research Paper
Abstract :Accurate estimation of the surface dose in radiotherapy is very important in reducing skin reactions. This study aims to evaluate the accuracy of two different treatment planning algorithms in calculating the surface dose in a specially designed phantom using thermoluminescent dosimetry insert ignore into journalissuearticles values(TLD);. In this study, a special phantom was designed for surface dose measurement. The phantom surface consisted of an adhesive bolus for the adhesion of TLDs. 121 TLDs were placed 1 cm apart on the bolus surface. In TPS, irradiation plans were created at different fields and source-surface distances insert ignore into journalissuearticles values(SSD);. Dose calculations were made with Anisotropic Algorithm algorithms insert ignore into journalissuearticles values(AAA); and Pencil Beam Convolution insert ignore into journalissuearticles values(PBC); algorithms for all plans. The mean dose was measured for each point. For each of the 4x4, 6x6, 8x8, 10x10, and 12x12 cm2 domains, the TLDs within the domain were approximately 1 cm inward from the edge. To measure the effect of SSD on surface dose, the isocenter point was located at depths of 0 cm, 2.5 cm and 5.0 cm, respectively. The surface dose at each depth was measured with TLDs. The doses calculated by the AAA and PBC algorithms were compared with the doses measured by TLDs. The AAA algorithm overestimates the surface dose by 4% compared to the TLD measurement for the 4x4 field. The surface dose calculation of the PBC algorithm was found to be high when compared to TLD measurements for all SSDs and fields. There was a significant difference between the PBC algorithm dose calculation and TLD measurements in all fields and SSDs insert ignore into journalissuearticles values(p<0.001);. It was observed that the AAA algorithm performed better in calculating the surface dose than the PBC algorithm. AAA and PBC algorithm users are advised to be more careful about surface dose calculation.Keywords : Surface Dose, Thermoluminescent Dosimetry, Algorithm