Effect of Water-Cooled Loop Monopole Antenna on Microwave Ablation Efficiency
Authors : Olumuyiwa ADEWUMİ, Okechukwu AGBASİ, Oluwanifesii FATİLE
Pages : 41-54
Doi:10.24998/maeusabed.590456
View : 11 | Download : 7
Publication Date : 2019-12-30
Article Type : Research Paper
Abstract :This work compares the effect of a monopole antenna without cooling unit on a biological tissue and a monopole antenna with cooling unit. Computer simulation was done using a computer software named COMSOL Multiphysics TM , input power was varied from 10W to 120W, with an interval of 10W, at different times, from 60s to 120s at an interval of 60s. It was observed that the diameter of the lowest power insert ignore into journalissuearticles values(20W); decreased by -17.2% at 5mins, the length decreased by -32.9% and aspect ratio for the two antennas increased by 23.40%. For medium power insert ignore into journalissuearticles values(60W);, at 5mins, the diameter decreased by -9.89%, length decreased by -37.9%, and aspect ratio decreased by 45.24%. The p-value obtained from diameter for low power showed no significant difference for ablation without water cool insert ignore into journalissuearticles values(p > 0.005);, but showed significant difference with water cool insert ignore into journalissuearticles values(p < 0.005);, diameter for medium power showed no significant difference, p > 0.005 in both simulations, there was no statistically significant difference in the diameter for high power insert ignore into journalissuearticles values(110W and 120W); for both simulations, p-value > 0.005. The result showed that inclusion of cooling unit reduced backward heating of microwave antenna, ablation length, and ablation diameter of simulated tissue.Keywords : Aspect ratio, Closed loop, Microwave ablation, Significant difference